InterLab Study
Our InterLab Study Lab Book can be found here for a fully detailed overview of our study.
Devices
The following devices were created in a backbone pSB1C3: ⚫ J23101 + I13504 (B0034-E0040-B0015) ⚫ J23106 + I13504 (B0034-E0040-B0015). However, constructing a device J23117 + I13504 (B0034-E0040-B0015) wasn't successful due to the time limit and limited ligation times. More details available on Lab Book. The device sizes were analyzed with restriction mapping which results can be seen in Figure 1.Used BBa_I20270, a GFP expressing part in the pSB1C3 backbone as a positive control. Negative controls were organisms with an empty plasmid and without any vector.
Protocols
Constructing devices
Protocols for the restriction, ligation and transformation of BioBricks can be found here.
Cultivations and measurement
Followed the next protocol for preparing the samples:Streaked out LB-plates with E.coli TOP10 strain containing every device and control with chloramphenicol concentration of 35ug/ml. Incubated plates overnight (18-20 hours) at 37C. Picked up biological triplates from the plates and inoculated 3 ml liquid cultures from the colonies with 12ml polypropylene test tube. Incubated the tubes at 37C with shaking at 300rpm for 18 hours. Measured OD600 values of cultures in cuvettes and calculated the dilution required for each sample to obtain the OD value of 0,5. Diluted and re-measured the samples to reach 5% accuracy of 0,5. Transfered 200ul of each sample into 96-well transparent plate and set the platereader to read fluorescence intensity. Proceed with the measurements with the following equipment: Type: Cell Imaging Multi-Mode Plate Reader
Name and model: Cytation 3
Company: BioTek Instruments, Inc
The following parametres were used:
Excitation wavelength: 470 nmEmission wavelength: 511 nmOptics Position: Bottom Filter: Quadruple Monochromators with Bandwith of 16 nm Units Reported: RFU
Results
Most samples generated overflow values which could not be measured with the plate reader. The incubation may have generated too much fluorescence proteins even though the samples were diluted properly. The problem wasn't solved after new dilutions of 9:10 ratio and due to the time limit, other ratios weren't attempted.
1st Technical Replicon
Sample |
J23117 + I13504 |
J23101 + I13504 |
J23151 |
Empty backbone |
Without backbone |
Blanks |
1 |
OVRFLW |
83954 |
53925 |
16309 |
13663 |
29336 |
2 |
OVRFLW |
OVRFLW |
56764 |
16335 |
14148 |
14940 |
3 |
OVRFLW |
OVRFLW |
61918 |
16771 |
14073 |
|
2nd Technical Replicon
Sample |
J23117 + I13504 |
J23101 + I13504 |
J23151 |
Empty backbone |
Without backbone |
Blanks |
1 |
OVRFLW |
85224 |
53553 |
16307 |
13747 |
29470 |
2 |
OVRFLW |
OVRFLW |
56472 |
16204 |
14267 |
14972 |
3 |
OVRFLW |
OVRFLW |
60969 |
16818 |
14381 |
|
3rd Technical Replicon
Sample |
J23117 + I13504 |
J23101 + I13504 |
J23151 |
Empty backbone |
Without backbone |
Blanks |
1 |
OVRFLW |
86750 |
56390 |
16123 |
13536 |
30778 |
2 |
OVRFLW |
OVRFLW |
58294 |
16294 |
14207 |
14856 |
3 |
OVRFLW |
OVRFLW |
62229 |
16793 |
13994 |
|
Restriction Map