Difference between revisions of "Team:KU Leuven/Research/Results"
Laetitia VW (Talk | contribs) |
Laetitia VW (Talk | contribs) |
||
Line 318: | Line 318: | ||
<div id=figure6>Figure 4</div> | <div id=figure6>Figure 4</div> | ||
First estimation of the OHHL standard curve. click to enlarge</h4> | First estimation of the OHHL standard curve. click to enlarge</h4> | ||
+ | </div> | ||
+ | </div> | ||
+ | |||
+ | <p>In the second experiment, a range between 0.01 and 0.10 mM was investigated. The difference with the previous experiment is that the OD was first measured in 1 cm cuvettes and that the violacein was afterwards isolated. In this way, the plate reader does not contain cells. This optimisation is done, because we noticed in the previous experiment that violacein was produced more quickly in a culture tube than in a microtiterplate, probably due to the amount of supplied oxygen. Another reason for this working method is because estimating the amount of cells is more standardised by using cuvettes of 1 cm than using a microtiterplate. | ||
+ | </p> | ||
+ | <p> Table 1 contains the processed values of the absorbance measurements. The processing is similar to the previous experiment. First the absorbance of DMSO was subtracted from the absorbance of the standards. Thereafter, these values were divided by the OD measured in 1 cm cuvettes. And finally, these values were corrected by putting the standard with a concentration of 0 mM OHHL in the origin. | ||
+ | </p> | ||
+ | |||
+ | <p> Here comes table 1 </p> | ||
+ | |||
+ | <p>In figure 5, our standard curve is plotted. A linear correlation between the absorbance and the concentration OHHL can be found. The variance of the technical replicates, visualised by the error bars, and the variance of the regression curve, shown by the R^2 value, can be explained by pipetting and measuring errors. Also, working with biological cells generates a background noise. This standard curve could give an estimation of bacterial AHL production. But it is important to keep in mind that there is background noise. Optimisation of this curve can be done by making more biological and technical replicas. | ||
+ | </p> | ||
+ | |||
+ | <div class ="center"> | ||
+ | <div id="imageAHL5"> | ||
+ | <a class="example-image-link" href="https://static.igem.org/mediawiki/2015/2/24/KU_Leuven_ResultOHHL4.jpeg" | ||
+ | data-lightbox="example-set" data-title="acceptor_cell"> | ||
+ | <img class="example-image" src="https://static.igem.org/mediawiki/2015/2/24/KU_Leuven_ResultOHHL4.jpeg" width="50%"></a> | ||
+ | <h4> | ||
+ | <div id=figure6>Figure 5</div> | ||
+ | Standard curve ranging from 0 to 0.1 mM. The error bars represent the standard deviation between the technical replicates.</h4> | ||
</div> | </div> | ||
</div> | </div> |
Revision as of 22:41, 18 September 2015
Results
Leucine detection
The standard curve from 0 to 100 µM did not give a linear relationship. Our working method needs optimisation. Because the enzymes are from other organisms than mentioned in Kugimiya and Fukada (2015), it is possible that the enzymes have another efficiency and as a consequence need to have another ratio (substrates over enzyme). Additionally, we did not have the same equipment as described in the article: we had to manually pipet the luminol solution. This possibly means that the measurements have a delay.
Due to a lack of time, we couldn’t complete the plasmid assembly and therefore, we were not able to proceed the quantification of leucine.
In comparison to HPLC, the chosen method would be less time consuming without the need of specialized equipment.
Contact
Address: Celestijnenlaan 200G room 00.08 - 3001 Heverlee
Telephone: +32(0)16 32 73 19
Email: igem@chem.kuleuven.be