Difference between revisions of "Team:Bielefeld-CeBiTec/Results/HeavyMetals"

Line 213: Line 213:
 
     <p>In addition, we performed an experiment in which the arsenic repressor was not present in the reaction from the beginning, but was encoded on a second plasmid. The plasmid concentrations we used had been predicted by our model. In accordance with the aforementioned results, we observed no clear repression and addition of arsenic showed no effect. This experiment is discussed on the <a href="https://2015.igem.org/Team:Bielefeld-CeBiTec/Modeling/Application">Modeling pages</a>.</p>
 
     <p>In addition, we performed an experiment in which the arsenic repressor was not present in the reaction from the beginning, but was encoded on a second plasmid. The plasmid concentrations we used had been predicted by our model. In accordance with the aforementioned results, we observed no clear repression and addition of arsenic showed no effect. This experiment is discussed on the <a href="https://2015.igem.org/Team:Bielefeld-CeBiTec/Modeling/Application">Modeling pages</a>.</p>
  
<h2>To sum it up</h2>
+
<h2>To summarize</h2>
 
<p>The arsenic sensor is the most prominent example of heavy metal biosensors. Our results confirm that it is possible to detect the safety limit of 10 µg/L arsenic with a cell-based arsenic senosor. The results of the tests in our CFPS indicate that this approach is suitable for the detection of arsenic as well. However, we believe that the genetic construct requires further optimization.</p>
 
<p>The arsenic sensor is the most prominent example of heavy metal biosensors. Our results confirm that it is possible to detect the safety limit of 10 µg/L arsenic with a cell-based arsenic senosor. The results of the tests in our CFPS indicate that this approach is suitable for the detection of arsenic as well. However, we believe that the genetic construct requires further optimization.</p>
  
Line 343: Line 343:
 
</div>
 
</div>
  
<h2>To sum it up<h2>
+
<h2>summarize<h2>
 
<p>Our chromium sensor detects the presence of chromium <i>in vivo</i>, but the outcome differed from our expectations. We would have expected an increase in fluorescence by increasing chromium concentrations. Our <i>in vitro</i> data suggest that these decrease in fluorescence could be explained by chromium’s influence on <i>E. coli</i> which is not reflected in growth but shown by chromium´s influence on the cell extract. Before normalizing the <i>in vitro</i> data the same pattern as <i>in vivo</i> could be observed. After normalization an increase in signal is noticeable. Therefore with optimization our chromium sensor would be compatible to our cell free sensor system.</p>
 
<p>Our chromium sensor detects the presence of chromium <i>in vivo</i>, but the outcome differed from our expectations. We would have expected an increase in fluorescence by increasing chromium concentrations. Our <i>in vitro</i> data suggest that these decrease in fluorescence could be explained by chromium’s influence on <i>E. coli</i> which is not reflected in growth but shown by chromium´s influence on the cell extract. Before normalizing the <i>in vitro</i> data the same pattern as <i>in vivo</i> could be observed. After normalization an increase in signal is noticeable. Therefore with optimization our chromium sensor would be compatible to our cell free sensor system.</p>
 
</div>
 
</div>
Line 467: Line 467:
  
  
<h2>To sum it up</h2>
+
<h2>To summarize</h2>
 
<p>Our copper sensors <i>in vivo</i> data show that detection of different copper concentrations is possible. The fluorescence levels defer clearly between different induction concentrations. As shown above higher copper concentration, higher the fluorescence signal. Therefore the concept of our sensor is functional even if the concentration needed for induction are to high to reach sensitively concerning the WHO guidelines for copper. Our sensor has been tested <i>in vitro</i> as well. For copper we tested our original CopAP construct without a T7 promoter in front of the inducible at first.  After realizing that the sensor shows the right tendencies but the general fluorescence is quite low we created an inducible promoter under the control of a T7 promoter to use in CFPS. Fluorescence levels of this device showed the same tendencies as the one without but were higher fluorescence’s, which helps detecting it.</p>
 
<p>Our copper sensors <i>in vivo</i> data show that detection of different copper concentrations is possible. The fluorescence levels defer clearly between different induction concentrations. As shown above higher copper concentration, higher the fluorescence signal. Therefore the concept of our sensor is functional even if the concentration needed for induction are to high to reach sensitively concerning the WHO guidelines for copper. Our sensor has been tested <i>in vitro</i> as well. For copper we tested our original CopAP construct without a T7 promoter in front of the inducible at first.  After realizing that the sensor shows the right tendencies but the general fluorescence is quite low we created an inducible promoter under the control of a T7 promoter to use in CFPS. Fluorescence levels of this device showed the same tendencies as the one without but were higher fluorescence’s, which helps detecting it.</p>
 
</div>
 
</div>
Line 505: Line 505:
 
The differences between inductions with various lead concentrations are really slight therefore this sensor needs further optimization which was not possible in this limited time. But as there is a fluorescence response to lead this sensor has the potential work as expected. In the future a characterization in CFPS systems would be interesting. </p>
 
The differences between inductions with various lead concentrations are really slight therefore this sensor needs further optimization which was not possible in this limited time. But as there is a fluorescence response to lead this sensor has the potential work as expected. In the future a characterization in CFPS systems would be interesting. </p>
  
<h2>To sum it up</h2>
+
<h2>To summarize</h2>
 
<p>Our lead sensor was characterized <i>in vivo</i> only. The differences between inductions with various lead concentrations are really slight therefore this sensor needs further optimization which was not possible in this limited time. But as there is a fluorescence response to lead this sensor has the potential work as expected. In the future a characterization in CFPS systems would be interesting.</p>
 
<p>Our lead sensor was characterized <i>in vivo</i> only. The differences between inductions with various lead concentrations are really slight therefore this sensor needs further optimization which was not possible in this limited time. But as there is a fluorescence response to lead this sensor has the potential work as expected. In the future a characterization in CFPS systems would be interesting.</p>
 
</div>
 
</div>
Line 609: Line 609:
  
  
<h2>To sum it up</h2>
+
<h2>To summarize</h2>
 
<p>Our mercury sensor works well <i>in vivo</i> as data show. There is a clearly noticeable increase in fluorescence after induction with mercury. Even the WHO guideline is measurable. This well working sensor was tested in <i>in vitro</i>as well. Taken data suggest, that maximal output is reached at concentrations of 6µg/L, which represent the former mentioned WHO guideline. With optimization a detection of even lower concentrations could be possible <i>in vitro</i>.</p>
 
<p>Our mercury sensor works well <i>in vivo</i> as data show. There is a clearly noticeable increase in fluorescence after induction with mercury. Even the WHO guideline is measurable. This well working sensor was tested in <i>in vitro</i>as well. Taken data suggest, that maximal output is reached at concentrations of 6µg/L, which represent the former mentioned WHO guideline. With optimization a detection of even lower concentrations could be possible <i>in vitro</i>.</p>
 
</div>
 
</div>
Line 645: Line 645:
 
With this sensor no production of sfGFP via fluorescence level change could be detected. Therefore this sensor is not suitable for approach. Therefore no <i> in vitro </i> data using CFPS were taken.</p></br>
 
With this sensor no production of sfGFP via fluorescence level change could be detected. Therefore this sensor is not suitable for approach. Therefore no <i> in vitro </i> data using CFPS were taken.</p></br>
  
<h2>To sum it up</h2>
+
<h2>To summarize</h2>
 
<p>With this sensor no production of sfGFP via fluorescence level change could be detected. Therefore, this sensor is not suitable for our approach. Consequently, no <i>in vitro</i> tests were performed.
 
<p>With this sensor no production of sfGFP via fluorescence level change could be detected. Therefore, this sensor is not suitable for our approach. Consequently, no <i>in vitro</i> tests were performed.
 
To create a working sensor based on this concept further optimization is needed.</p>
 
To create a working sensor based on this concept further optimization is needed.</p>

Revision as of 16:26, 18 September 2015

iGEM Bielefeld 2015


Heavy Metals

Results

Adjusting the detection limit
Influence of heavy metals on the growth of E.coli KRX. The tested concentrations were 20 µg/L lead, 60 µg/L mercury, 60 µg/L chromium, 80 µg/L nickel, 40 mg/L copper, which represent ten times the WHO guideline. The influence of arsenic was not tested as E. coli is known to be resistant to arsenic.

We tested our heavy metal biosensors in Escherichia coli as well as in our cell-free protein synthesis.

Prior to the in vivo characterization, we tested whether the heavy metals have a negative effect on the growth of E. coli.

As can be seen from the figure, we observed no significant difference between the growth in the presence of heavy metals and the controls. This first experiment showed us that in vivo characterization of these sensors is possible. Most cultivations for in vivo characterization were performed in the BioLector. Due to the accuracy of this device, we could measure our samples in duplicates. Subsequently, all functional biosensors were tested in vitro.

Click on the test strip for the results of our biosensor tests in E. coli and in our CFPS:

teststrip

To sum it all up

We have characterized heavy metal sensors for arsenic, chromium, copper, lead, mercury and nickel. The results for our nickel characterization indicated that the constructed nickel sensor is not suitable for our test strip. The sensors for lead and chromium showed great potential, as they showed responses to chromium or lead, but require further optimization. Copper, our new heavy metal sensor, worked as expected and was able to detect different copper concentrations. The already well-characterized sensors for arsenic and mercury were tested as well. While the arsenic sensor worked well in vivo, it requires some omptimization for the use in vitro. Mercury showed that a fully optimized sensor works very well in our in vitro system and has the potential to detect even lower concentrations than in vivo.