Difference between revisions of "Team:Tokyo Tech/Experiment/Interlab"
Line 94: | Line 94: | ||
</tr> | </tr> | ||
</table><br> | </table><br> | ||
+ | <p class="text2">We measured three colonies(#1〜3) three times (Technical replicate 1〜3) per each sample (Device1〜3, positive control and negative control).<br> | ||
+ | Using the calibration curve (Figure. 3-7-3-1), we were able to obtain the absolute unit of fluorescence (Table. 3-7-3-1). The way we obtained the absolute unit is described in the 4. Material and Method section. These results show that the intensity of fluorescence was in the following order, Device1>Device2>positive control>Device3>negative control.</p> | ||
+ | <table width="940 px" border="0px"> | ||
+ | <tr> | ||
+ | <td width="940px"><div align="center"><img src="https://static.igem.org/mediawiki/2015/c/ca/Tokyo_Tech_Interlab_Table.3-7-3-1.png" width="700px"/> | ||
+ | </td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td width="940px"> | ||
+ | <h4 align="center" class="fig"><strong>Table. 3-7-3-1.</strong> The absolute unit of fluorescence intensity</h4> | ||
+ | <td> | ||
+ | </tr> | ||
+ | </table><br> | ||
+ | |||
+ | |||
+ | |||
+ | |||
<h2 id="Discussion" class="smalltitle">4. Discussion</h2> | <h2 id="Discussion" class="smalltitle">4. Discussion</h2> | ||
まだまだいける。 | まだまだいける。 |
Revision as of 19:25, 12 September 2015
Interlab
We have measured three devices.
contents
1. Introduction
2. Summary of the Experiment
3. Results
3-1. Plate reader
3-2. Plasmid
5-3. Assay Protocol
4. Discussion
5. Materials and Methods
5-1. Strains
5-2. Plasmid
5-3. Assay Protocol
6. Reference
1. Introduction
In iGEM 2015, the Interlab Study was held, where we measured the expression level of GFP using three designated devices. It was the first time for our team to join this Interlab Study. In addition to the three designated devices, we also measured the expression level of GFP from a positive control and a negative control using the flow cytometer and the plate reader. Also, for the plate reader, we succeeded in calculating the absolute unit by drawing the calibration curve using sodium fluorescein.
2. Summary of the Experiment
Our purpose was to obtain the fluorescence data of the three designated devices and to compare them. We prepared Device1〜Device3, Positive control and Negative control as shown below. We measured the exact same colonies of the exact same samples with both the plate reader and the flow cytometer.
Device 1: J23101 + I13504(pSB1C3)
Device 2: J23106 + I13504(pSB1C3)
Device 3: J23117 + I13504(pSB1C3)
Positive control: BBa_I20270(pSB1C3)
Negative control: BBa_R0040(pSB1C3)
Fig.3-7-2-1. designated devices |
3. Results
3-1. Plate reader
First of all, we calibrated our plate reader by confirming the linear relationship between sodium fluorescein concentration and fluorescence (Figure. 3-7-3-1). The way we obtained the calibration curve is descried in the 4. Material and Method section.
Fig.3-7-3-1. Calibration curve |
We measured three colonies(#1〜3) three times (Technical replicate 1〜3) per each sample (Device1〜3, positive control and negative control).
Using the calibration curve (Figure. 3-7-3-1), we were able to obtain the absolute unit of fluorescence (Table. 3-7-3-1). The way we obtained the absolute unit is described in the 4. Material and Method section. These results show that the intensity of fluorescence was in the following order, Device1>Device2>positive control>Device3>negative control.
Table. 3-7-3-1. The absolute unit of fluorescence intensity |
We measured three colonies(#1〜3) three times (Technical replicate 1〜3) per each sample (Device1〜3, positive control and negative control).
Using the calibration curve (Figure. 3-7-3-1), we were able to obtain the absolute unit of fluorescence (Table. 3-7-3-1). The way we obtained the absolute unit is described in the 4. Material and Method section. These results show that the intensity of fluorescence was in the following order, Device1>Device2>positive control>Device3>negative control.
Table. 3-7-3-1. The absolute unit of fluorescence intensity |