Revision as of 01:18, 19 September 2015
- Project
- Modeling
- Lab
- Human
Practices - Parts
- About Us
InterLab Study
Introduction
Since 2014, teams participating in iGEM have been invited to collaborate in the InterLab study, with the objective of comparing the variability of measurements among different research groups.
Goal
The participating labs were tasked with measuring the fluorescence of three different devices consisting of three promoters (strong, medium and weak) followed by a GFP gene, with the aim of comparing the results among different iGEM teams all around the world. The recommended standard method for measuring is a plate reader setup. However, teams were invited to use other methods. In our case, we measured the devices using the plate reader and FACS.
Methodology
Protocol
For this experiment, we followed the protocol available in the iGEM website.
Organisms were prepared by BioBrick assembly using the parts indicated in Table 1. Once the parts were ligated, the plasmids containing devices 1-3 and the positive and negative controls were transformed into TOP10 Escherichia coli. The three devices, the positive control (I20270), and the first negative control (B0012) were cultured on LB plates with the addition of chloramphenicol, while TOP10 cells without the addition of any plasmid were cultured in LB-Streptomycin plates. The transformed cells were incubated in this manner overnight for 16 h at 37ºC.
Table 1. Summary of used elements
Devices | Device 1 | J23101 + I13504 |
---|---|---|
Device 2 | J23106 + I13504 | |
Device 3 | J23117 + I13504 | |
Controls | Positive control | I20270 |
Negative control | no plasmid | |
B0012 |
Ten colonies were picked from each plate. The presence of the correct device in the colonies was checked first by colony PCR and, if the size of the DNA fragment corresponded to the expected size, were then sent for sequencing. The colonies containing the correct plasmid were then incubated overnight for 16 h at 37ºC in 12 mL conical flasks in 5 mL LB medium with chloramphenicol or streptomycin, respectively. The flasks were shaken at 200 rpm at an angle.
Protocol for the plate reader
The overnight bacteria culture was diluted in Lysogenic Broth (LB) to OD600 of 0.2±0.04 in a 96-well plate. OD600 was monitored during the measurement for verification of the concentration. A unique measurement of the devices’ fluorescence was performed at an excitation wavelength of 488 nm and emission was monitored at a wavelength of 530 nm 10 min after the dilution.
Protocol for FACS
Overnight cultures in LB were diluted 1:10 before the measurement. FACS was calibrated with PBS filtered with a 0.2 µm filter. Samples were diluted to have no more than 4,000 events/s prior to the measurement. TOP10 cells without the plasmid were used to establish the gating for GFP positive and negative cells, respectively. For the measurement of the three devices and all controls, 100,000 events were recorded in biological, as well as technical triplicates.
Machines
Incubator
The incubator used for growing the cells was the KuhnerShakerX, model ClimoShaker ISF1-X. The throw was orbital with 25mm of diameter.
FACS
The FACS machine used was a BD LSRFortessa Cell Analyzer. Cells were excited at a wavelength of 488nm and the filter used was a 505 LP. The SSC threshold was 400. Data is presented in absolute units.
The last calibration before our use was done on the 7th July 2015 by Ms Jäggin.
Data was analyzed using Diva, FlowJo 10.0.8 and Microsoft Excel 2013.
Plate reader
The plate reader we used is the Tecan Infinite M200 Pro. The excitation wavelength was set to 488 nm and the emission was at 530 nm. Fluorescence was normalized with the optical density, using fluorescence units/absorbance units. Optical density was determined at 600 nm.
The last calibration of the plate reader was done on the 17th March 2015 by the official Tecan Service (Tecan Sales Switzerland AG).
Data was analysed using Microsoft Excel 2013.
Results
FACS measurement
Figure 1. A) Comparison in the fluorescence of TOP10 cells without plasmid (black), with B0012 (light grey), with I20270 (dark grey), device 1 (orange), device 2 (blue) and device 3 (red). B) Mean values of fluorescence for each device. n=3 ± SD
The median value was calculated for FACS as the distribution was skewed. We measured three different colonies (biological replicates) three times each (technical replicates). Data for each of them are summarized in Table 2.
Plate reader
As can be seen from Table 2 and Figure 2, the data correspond to the expected results: almost no fluorescence is detected in controls with TOP10 and TOP10 with B0012 plasmid, whereas cells transformed with I20270 present intermediate fluorescence. Device 1 has the strongest fluorescence, followed by device 2, and finally device 3 which is the weakest.
Other data
Who participated in the study?
Lisa Baumgartner, Anna Fomitcheva, Anja Michel and Michael Meier prepared the transformants.
Verena Jäggin showed us how to properly manipulate the FACS machine and use the software.
Lisa Baumgartner and Anna Fomitcheva did the readings with FACS.
Lisa Baumgartner and Michael Meier did the plate reader experiment.
Data was processed by Lisa Baumgartner, Anna Fomitcheva and Michael Meier.
Dates in which the study took place
The measurements for the InterLab study took place on the 10th July 2015 for FACS and on 21st July 2015 for the plate reader.