Team:Freiburg/Protocols/Slide Wash
iRIf slide wash steps
iRIf slide wash steps after measurement/before new measurement
remarks concerning the protocol (author, adapted from etc.)
material: 0,1% SDS, 10% RBS, diH2O, Kimtechtuch
duration: Incubation in SDS o/n + ~10 mins for further washing
- Add volume of added chemical (µL) (remarks, tips)
- instructions time, speed, temperature (remarks, tips)
After measuring with iRIf device:
- Fill slide-box with 0,1% SDS; incubate Slides o/n
- rub slide surface with 0,1% SDS with Kimtechtuch
- wash with diH2O from wash bottle
- dry slides with cleaning wipe/nitrogenpistole
- incubate in 10%ige RBS (glasbox)for 1-3 minutes; twist slides 1x within this time window
- rub slide surface with 10% RBS with Kimtechtuch (RBS that is already on the slides is enough)
- wash properly with diH2O
- dry with nitrogenpistole
- when completely dry (change position of the pincette at least one time to get rid of the water that is covered through it) sort slides into slidebox
- plasma activate slides for 5 min
- store slides in iRIf slide box
- Wash pincettes before and after use with aceton (Caution: do not wash them in top of the H2O/Ethanol-waste, seperate waste!)
- Slide „303“ is not properly cleanable, avoid using it