Part | Description | Manufacturer / Partname | Price[USD] |
---|---|---|---|
A fairly strong LED | In our case a 3W green 520 nm LED | Cree XP-E on star circuit board | 10 |
A cooling element | Used to prevent the LED from overheating | Fischer Elektronik ICK S | 9 |
A constant-current-source | To ensure static LED light (no flickering) | Roschwege GmbH KSQ-3W | 15 |
An AC/DC rectifier | Used to grant direct current to the LED | Goobay 67951 | 16 |
Two optical lenses | Here we used two identical 60 mm lenses | Thorlabs AC254-060-A | 160 |
A camera | We used a SLR camera | Canon 550D | 200-400 |
A microfluidics chamber | Our chamber consists of an iRIf slide attached to a PDMS flowcell | Made ourselves | |
Magnets | Used to attach the flowcell to the device | 5 mm Neodymium magnets | 3 |
A small syringe | We used a regular 10 ml syringe | BRAUN INJEKT 10 ml | < 1 |
A microfluidic tube | Used to connect the flowchamber with the syringe | - | < 1 |
A case for the device | 5 mm thick acrylic glass cutouts to hold the parts in place | Ordered at http://formulor.com | 100 |
Difference between revisions of "Team:Freiburg/Results/Own Device"
Line 71: | Line 71: | ||
<p> | <p> | ||
− | For testing our device we immobilized proteins derived from rabbits on an iRIf slide in distinct spots (figure 1 D). The proteins we used were polyclonal anti-HCV (Hepatitis C Virus) antibodies derived from rabbit, which we then aimed to detect with anti-rabbit antibodies. We used these interaction partners in this series of | + | For testing our device we immobilized proteins derived from rabbits on an iRIf slide in distinct spots (figure 1 D). The proteins we used were polyclonal anti-HCV (Hepatitis C Virus) antibodies derived from rabbit, which we then aimed to detect with anti-rabbit antibodies. We used these interaction partners in this series of experiments to reproduce a successful measurement, which we previously performed in the professional measuring device. |
The binding layer on the iRIf slide consisted of an APTES/PDITC surface. The spots on the slide were produced by pipetting 3 µl [500 µg/ml] rabbit-anti-HCV protein and 3 µl [1 mg/ml] BSA in an alternating pattern onto the slide. After incubation the slide was blocked for 30 min in BSA solution. | The binding layer on the iRIf slide consisted of an APTES/PDITC surface. The spots on the slide were produced by pipetting 3 µl [500 µg/ml] rabbit-anti-HCV protein and 3 µl [1 mg/ml] BSA in an alternating pattern onto the slide. After incubation the slide was blocked for 30 min in BSA solution. | ||
</p> | </p> | ||
Line 85: | Line 85: | ||
</div> | </div> | ||
<p> | <p> | ||
− | The camera used for the measurement was a Canon 50D. The camera was set to automatically take pictures at an interval of | + | The camera used for the measurement was a Canon 50D. The camera was set to automatically take pictures at an interval of five seconds. The exposure time was set in order for the pixels in the image to have approx. 80% of maximum light saturation before the solution was flushed onto the chip (figure 1 C & D). |
− | + | The antibody solution was pipetted into the flowchamber without the use of any microfluidic pump. Instead, a syringe was loaded with 500 µl [5 µg/ml] anti-rabbit antibody solution (diluted in PBS) and slowly released into the binding chamber of the device by gently dispensing it from the syringe. | |
As can be seen in figure 1 C, the quotient picture clearly shows binding of anti-rabbit antibodies to the rabbit protein spots. The BSA control spots show none or negligible unspecific binding. | As can be seen in figure 1 C, the quotient picture clearly shows binding of anti-rabbit antibodies to the rabbit protein spots. The BSA control spots show none or negligible unspecific binding. | ||
Line 166: | Line 166: | ||
</tr> | </tr> | ||
<tr class="row7"> | <tr class="row7"> | ||
− | <td class="col0">A microfluidics chamber</td><td class="col1">Our chamber consists of an iRIf slide attached to a PDMS | + | <td class="col0">A microfluidics chamber</td><td class="col1">Our chamber consists of an iRIf slide attached to a PDMS flowcell</td><td class="col2">Made ourselves</td><td class="col3"></td> |
</tr> | </tr> | ||
<tr class="row8"> | <tr class="row8"> | ||
− | <td class="col0">Magnets</td><td class="col1">Used to attach the | + | <td class="col0">Magnets</td><td class="col1">Used to attach the flowcell to the device</td><td class="col2">5 mm Neodymium magnets</td><td class="col3">3</td> |
</tr> | </tr> | ||
<tr class="row9"> | <tr class="row9"> | ||
Line 175: | Line 175: | ||
</tr> | </tr> | ||
<tr class="row10"> | <tr class="row10"> | ||
− | <td class="col0">A microfluidic tube</td><td class="col1"> Used to connect the | + | <td class="col0">A microfluidic tube</td><td class="col1"> Used to connect the flowchamber with the syringe</td><td class="col2">-</td><td class="col3">< 1</td> |
</tr> | </tr> | ||
<tr class="row11"> | <tr class="row11"> | ||
Line 199: | Line 199: | ||
− | <div class="thumb2 trien" style="width:410px;padding-top:10px"><div class="thumbinner"><a class="media lightbox_trigger" href="https://static.igem.org/mediawiki/2015/8/88/Freiburg_results-device_3d_no_walls.png" title="results:device_3d_no_walls.png"><img alt="" class="mediabox2" src="https://static.igem.org/mediawiki/2015/8/88/Freiburg_results-device_3d_no_walls.png" width="400"/></a><div class="thumbcaption"><div class="magnify"><a class="internal" href="https://static.igem.org/mediawiki/2015/8/88/Freiburg_results-device_3d_no_walls.png" title="vergrößern"><img alt="" height="11" src="/igem2015/lib/plugins/imagebox/magnify-clip.png" width="15"/></a></div><strong>Figure 5: The 3D model of our device without walls and top part.</strong> Parts have been colorized for clarification - Green: The wall where light exits the device and enters the camera; Pink: a platform holding Cooling-Element+LED in place; Blue: platforms for holding the lenses in place; Gray: rear end wall where the | + | <div class="thumb2 trien" style="width:410px;padding-top:10px"><div class="thumbinner"><a class="media lightbox_trigger" href="https://static.igem.org/mediawiki/2015/8/88/Freiburg_results-device_3d_no_walls.png" title="results:device_3d_no_walls.png"><img alt="" class="mediabox2" src="https://static.igem.org/mediawiki/2015/8/88/Freiburg_results-device_3d_no_walls.png" width="400"/></a><div class="thumbcaption"><div class="magnify"><a class="internal" href="https://static.igem.org/mediawiki/2015/8/88/Freiburg_results-device_3d_no_walls.png" title="vergrößern"><img alt="" height="11" src="/igem2015/lib/plugins/imagebox/magnify-clip.png" width="15"/></a></div><strong>Figure 5: The 3D model of our device without walls and top part.</strong> Parts have been colorized for clarification - Green: The wall where light exits the device and enters the camera; Pink: a platform holding Cooling-Element+LED in place; Blue: platforms for holding the lenses in place; Gray: rear end wall where the flowchamber is attached to. The slits in the top and bottom part where the magnets have to be fixed</div></div></div> |
</div> | </div> | ||
Line 215: | Line 215: | ||
<a href="https://static.igem.org/mediawiki/2015/4/4c/Freiburg_results-device_parts_vector.png" class="lightbox_trigger" title="results:device_parts_vector.png"><img alt="Freiburg iRIf device parts" src="https://static.igem.org/mediawiki/2015/4/4c/Freiburg_results-device_parts_vector.png" width="400"/></a> | <a href="https://static.igem.org/mediawiki/2015/4/4c/Freiburg_results-device_parts_vector.png" class="lightbox_trigger" title="results:device_parts_vector.png"><img alt="Freiburg iRIf device parts" src="https://static.igem.org/mediawiki/2015/4/4c/Freiburg_results-device_parts_vector.png" width="400"/></a> | ||
<div class="thumbcaption"> | <div class="thumbcaption"> | ||
− | <strong>Figure 6</strong>: An image of the vector file used to order the parts. The vector file is used to | + | <strong>Figure 6</strong>: An image of the vector file used to order the parts. The vector file is used to excise parts of a 5 mm acrylic glass board with a laser. Black lines are cut out by the laser, gray lines/areas are engravings. Refer to the <a href="#device_download_links">download section</a> to downloaded the vector graphic files.</div> |
</div> | </div> | ||
</div> | </div> | ||
</div> | </div> | ||
<p> | <p> | ||
− | The most difficult part to build by oneself is the PDMS | + | The most difficult part to build by oneself is the PDMS flowchamber. For producing such flowchambers, a wafer has to be made in a cleanroom. The template used for our PDMS flowcell is shown in figure 9. If your facility has an engineering department, there is a high chance that a cleanroom is present. Ask the personel in charge if they can help you out with producing your flowcell. If you already use another type of microfluidic flowchambers, our device should still be compatible with it, though some adaptions might be necessary. If you have no possibility of creating a flowchamber, you might want to consider building a very simplified DIY flowchambers from two glass slides, separated by thin duct tape. Note that our device was not primarily build to support such a chamber though and would need appropriate modifications. |
</p> | </p> | ||
Line 230: | Line 230: | ||
<a href="https://static.igem.org/mediawiki/2015/2/27/Freiburg_Device_Flowcell_chamber_syringe.jpg" class="lightbox_trigger" title="Flow cell attachment"><img alt="Freiburg iRIf device flow cell attachment" src="https://static.igem.org/mediawiki/2015/0/08/Freiburg_Device_Flowcell_chamber_syringe_preview.jpg" width="400"/></a> | <a href="https://static.igem.org/mediawiki/2015/2/27/Freiburg_Device_Flowcell_chamber_syringe.jpg" class="lightbox_trigger" title="Flow cell attachment"><img alt="Freiburg iRIf device flow cell attachment" src="https://static.igem.org/mediawiki/2015/0/08/Freiburg_Device_Flowcell_chamber_syringe_preview.jpg" width="400"/></a> | ||
<div class="thumbcaption"> | <div class="thumbcaption"> | ||
− | <strong>Figure 7</strong> - left: a flow chamber consisting of an iRIf slide attached to a PDMS | + | <strong>Figure 7</strong> - left: a flow chamber consisting of an iRIf slide attached to a PDMS flowcell - center: a syringe attached to a pipette tip, used to inject antibody solutions into the flowchamber - right: the part of our device which connects the flowchamber, the syringe and the iRIf device. |
</div> | </div> | ||
</div> | </div> | ||
Line 254: | Line 254: | ||
<a href="https://static.igem.org/mediawiki/2015/0/00/Freiburg_Masken_M25-M33-1.jpg" class="lightbox_trigger" title="PDMS flow chamber"><img alt="Freiburg iRIf device flow cell" src="https://static.igem.org/mediawiki/2015/0/00/Freiburg_Masken_M25-M33-1.jpg" width="400"/></a> | <a href="https://static.igem.org/mediawiki/2015/0/00/Freiburg_Masken_M25-M33-1.jpg" class="lightbox_trigger" title="PDMS flow chamber"><img alt="Freiburg iRIf device flow cell" src="https://static.igem.org/mediawiki/2015/0/00/Freiburg_Masken_M25-M33-1.jpg" width="400"/></a> | ||
<div class="thumbcaption"> | <div class="thumbcaption"> | ||
− | <strong>Figure 9:</strong> Layout used to produce the wafer of our PDMS | + | <strong>Figure 9:</strong> Layout used to produce the wafer of our PDMS flowcell. |
</div> | </div> | ||
</div> | </div> |
Revision as of 13:41, 17 September 2015
Results of Our Self-Built Device
Here you will see what we were able to measure with the very first prototype of our self-built device. To see the results of our final device, please refer to the essential results page. A detailed explanation on how to build your own low-priced iRIf device can be found on the bottom of this page.
Our First Prototype is Able to Detect Antigen-Antibody Binding
For testing our device we immobilized proteins derived from rabbits on an iRIf slide in distinct spots (figure 1 D). The proteins we used were polyclonal anti-HCV (Hepatitis C Virus) antibodies derived from rabbit, which we then aimed to detect with anti-rabbit antibodies. We used these interaction partners in this series of experiments to reproduce a successful measurement, which we previously performed in the professional measuring device. The binding layer on the iRIf slide consisted of an APTES/PDITC surface. The spots on the slide were produced by pipetting 3 µl [500 µg/ml] rabbit-anti-HCV protein and 3 µl [1 mg/ml] BSA in an alternating pattern onto the slide. After incubation the slide was blocked for 30 min in BSA solution.
The camera used for the measurement was a Canon 50D. The camera was set to automatically take pictures at an interval of five seconds. The exposure time was set in order for the pixels in the image to have approx. 80% of maximum light saturation before the solution was flushed onto the chip (figure 1 C & D). The antibody solution was pipetted into the flowchamber without the use of any microfluidic pump. Instead, a syringe was loaded with 500 µl [5 µg/ml] anti-rabbit antibody solution (diluted in PBS) and slowly released into the binding chamber of the device by gently dispensing it from the syringe. As can be seen in figure 1 C, the quotient picture clearly shows binding of anti-rabbit antibodies to the rabbit protein spots. The BSA control spots show none or negligible unspecific binding.
How To Build Your Own Device
In this section we will demonstrate how to build your own iRIf device from scratch, using affordable, low-tech material. The design is focused on creating a device that is both low-priced and portable, demonstrating the potential for the DiaCHIP to be used even in rural areas where high-tech laboratories are poorly accessible.
General Principle
As can be seen in figure 3, the basic setup is fairly simple. Light from an LED enters a lense to obtain a parallel light beam. To achieve this, the distance from the LED to the lense has to be exactly one focal length. The light hits the iRIf slide afterwards where it is reflected (in the same angle as it hits the slide) and enters a second lense, whose purpose is to project a sharp image of the slide onto the CCD chip of the camera.
Construction Guidance
A major problem that we confronted when building the device from scratch was to assure that all components are at the exact distance and angle to each other. This is crucial as a slight misplacement of a component may lead to lower signal strength, blurred images or, in the worst case, no signal at all. This can be difficult since our device does not rely on straight angles. We overcame this problem by designing a case for the device that ensures the right placement of the components inside the device. This was achieved by calculating all the distances beforehand using physical laws of optics. We realized this by drawing an exact vector graphic blueprint for our device. We then constructed a digital 3D model of the casing based on the vector blueprint to avoid a costly and time-consuming trial and error process (figure 4).
We designed the casing so that all the necessary parts (lenses, LED) are held in the correct position safely during the measurement, but remain removable to grant easy transportation of the device (i.e. to the Giant Jamboree). Figure 5 illustrates the parts that hold the lenses and LED in place.
After assuring that the 3D model of our device was set-up correctly, we created a vector graphic file which layed out all the parts needed for the casing in a 2D plain. Using this vector graphic file, we ordered the parts at Formulor, a service which lasers out parts from acrylic glass using a vector graphic as a template. The vector graphic template is shown in figure 6, and may be downloaded and used by everyone to build their own device. To allow easy mounting of the casing we also provide an easy to understand manual. One advantage of using acrylic glass is that the parts can be glued together easily using a few drops of acetone, fusing the parts together. The vector graphic file used to laser our the casing parts also contains a template for the parts which are necessary to attach the glass and PDMS slide to the device. These parts have to be cut from 1 mm thick acrylic glass. Once they are cut from the acrylic glass with a laser, the microfluidic tubes can be glued to the parts (figure 7). Once this is completed, these parts can be attached to the device with the magnets (figure 8)
The most difficult part to build by oneself is the PDMS flowchamber. For producing such flowchambers, a wafer has to be made in a cleanroom. The template used for our PDMS flowcell is shown in figure 9. If your facility has an engineering department, there is a high chance that a cleanroom is present. Ask the personel in charge if they can help you out with producing your flowcell. If you already use another type of microfluidic flowchambers, our device should still be compatible with it, though some adaptions might be necessary. If you have no possibility of creating a flowchamber, you might want to consider building a very simplified DIY flowchambers from two glass slides, separated by thin duct tape. Note that our device was not primarily build to support such a chamber though and would need appropriate modifications.
Download section
Legal notice
The iRIf detection method is patented. Biametrics and associated persons own patents concerning the detection principle. These patents are:
- Method for examining physical, chemical and biochemical interactions (DE102004051098.9, DE102005015030.6, EP05797776.1)
- Study of molecular interactions on and / or in thin layers (DE102007038797.2)
- Method and apparatus for determining reflection coefficients to filter arrangement with thin layers (DE102009019711.7)
- Again Detectable support for optical measuring methods (DE102009019476.2)