Difference between revisions of "Team:Glasgow/Interlab"
Line 493: | Line 493: | ||
In order to calibrate the spectrophotometer a dilution series of 1-100% of DH5 alpha cells was carried out and the A600 of each sample was measured (Figure 1). | In order to calibrate the spectrophotometer a dilution series of 1-100% of DH5 alpha cells was carried out and the A600 of each sample was measured (Figure 1). | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/2/24/2015_InterLab_13.png"> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure> <img src="https://static.igem.org/mediawiki/2015/2/24/2015_InterLab_13.png"> | ||
<figcaption>Figure 1: Spectrophotometer calibration curve | <figcaption>Figure 1: Spectrophotometer calibration curve | ||
</figcaption> | </figcaption> | ||
Line 506: | Line 508: | ||
A dilution series was measured for phiLOV protein (Figure 2), converted to numerical readings (Table 1) and a calibration curve (Figure 3) carried out to calibrate the Typhoon. Fluorescent proteins derived from voltage (LOV) domains are smaller and more efficient under anaerobic conditions than green fluorescent proteins (GFP) (Buckley et, al. 2015). iLOV, an improved LOV flavoprotein, was originally engineered as a reporter for viral infection from phototropin, the blue light receptor. We used phiLOV which is a photostable version of the iLOV fluorescence reporter. | A dilution series was measured for phiLOV protein (Figure 2), converted to numerical readings (Table 1) and a calibration curve (Figure 3) carried out to calibrate the Typhoon. Fluorescent proteins derived from voltage (LOV) domains are smaller and more efficient under anaerobic conditions than green fluorescent proteins (GFP) (Buckley et, al. 2015). iLOV, an improved LOV flavoprotein, was originally engineered as a reporter for viral infection from phototropin, the blue light receptor. We used phiLOV which is a photostable version of the iLOV fluorescence reporter. | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/6/62/2015-Glasgow-interlab1.png"> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure> <img src="https://static.igem.org/mediawiki/2015/6/62/2015-Glasgow-interlab1.png"> | ||
+ | |||
<figcaption>Figure 2: Fluorescence readings of a dilution series of phiLOV. 67.5µg = 67.5µg phiLOV in 100µl PBS. Each concentration was carried out twice. | <figcaption>Figure 2: Fluorescence readings of a dilution series of phiLOV. 67.5µg = 67.5µg phiLOV in 100µl PBS. Each concentration was carried out twice. | ||
</figcaption> | </figcaption> | ||
Line 530: | Line 535: | ||
The devices, as shown in Table 2, were prepared using BioBrick assembly. Parts J23101, J23106, J23117, I13504, I20270 and R0040 were taken from the iGEM distribution plates and each transformed into TOP-10 competent cells. The promoters were digested with Pst1 and Spe1 and the GFP part, I13504, was digested with Xba1 and Pst1. The I13504 part was then ligated into each promoter plasmid and transformed into TOP-10 cells to create the three required devices in pSB1C3 (Figure 4). Restreaks were carried out for one colony of each device and control and three colonies of each (labelled 1, 2 and 3) were picked and grown separately. Sequencing was carried out to check the correct devices had been created. | The devices, as shown in Table 2, were prepared using BioBrick assembly. Parts J23101, J23106, J23117, I13504, I20270 and R0040 were taken from the iGEM distribution plates and each transformed into TOP-10 competent cells. The promoters were digested with Pst1 and Spe1 and the GFP part, I13504, was digested with Xba1 and Pst1. The I13504 part was then ligated into each promoter plasmid and transformed into TOP-10 cells to create the three required devices in pSB1C3 (Figure 4). Restreaks were carried out for one colony of each device and control and three colonies of each (labelled 1, 2 and 3) were picked and grown separately. Sequencing was carried out to check the correct devices had been created. | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/7/70/2015-Glasgow-interlab4.jpg" height="40%" width="40%"/> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure> <img src="https://static.igem.org/mediawiki/2015/7/70/2015-Glasgow-interlab4.jpg" height="40%" width="40%"/> | ||
<figcaption>Table 2: Summary of BioBrick used</figcaption> | <figcaption>Table 2: Summary of BioBrick used</figcaption> | ||
</br> | </br> | ||
Line 571: | Line 578: | ||
A dilution series of FAM labelled oligonucleotide was measured (Figure 5) and converted to numerical readings (Table 3) to enable absolute values for the devices to be calculated. The calibration curve (Figure 6) has a line gradient of 4.79x10^6. Therefore the fluorescence readings of the devices will be divided by the conversion factor of 4,790,000 to give absolute fluorescence as equivalent to pmol of FAM labelled oligonucleotide. Absolute values should be comparable across different equipment and protocols. | A dilution series of FAM labelled oligonucleotide was measured (Figure 5) and converted to numerical readings (Table 3) to enable absolute values for the devices to be calculated. The calibration curve (Figure 6) has a line gradient of 4.79x10^6. Therefore the fluorescence readings of the devices will be divided by the conversion factor of 4,790,000 to give absolute fluorescence as equivalent to pmol of FAM labelled oligonucleotide. Absolute values should be comparable across different equipment and protocols. | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/8/8d/2015-Glasgow-interlab6.png"> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure><img src="https://static.igem.org/mediawiki/2015/8/8d/2015-Glasgow-interlab6.png"> | ||
<figcaption>Figure 5: Fluorescence readings of a dilution series of FAM labelled oligonucleotide. 10pmol = 10pmol FAM labelled oligonucleotide in 100µl PBS.</figcaption> | <figcaption>Figure 5: Fluorescence readings of a dilution series of FAM labelled oligonucleotide. 10pmol = 10pmol FAM labelled oligonucleotide in 100µl PBS.</figcaption> | ||
</br> | </br> | ||
Line 592: | Line 601: | ||
The A600 of each device colony 1-3 and technical replicates were measured along with the controls (table 4). | The A600 of each device colony 1-3 and technical replicates were measured along with the controls (table 4). | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/d/d0/2015-Glasgow-interlab9.png"> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure> <img src="https://static.igem.org/mediawiki/2015/d/d0/2015-Glasgow-interlab9.png"> | ||
<figcaption>Table 4: Absorbance at 600nm for each biological and technical replicates of the devices and controls. Units are arbitrary.</figcaption> | <figcaption>Table 4: Absorbance at 600nm for each biological and technical replicates of the devices and controls. Units are arbitrary.</figcaption> | ||
</br> | </br> | ||
Line 618: | Line 629: | ||
4. Dividing these values by the conversion factor as determined from the FAM oligo dilutions (479,000) gives the absolute fluorescence equivalent to pmol of FAM oligo per A600 of cells (Table 6). | 4. Dividing these values by the conversion factor as determined from the FAM oligo dilutions (479,000) gives the absolute fluorescence equivalent to pmol of FAM oligo per A600 of cells (Table 6). | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/f/ff/2015-Glasgow-interlab21.png" height="70%" width="70%"/> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure> <img src="https://static.igem.org/mediawiki/2015/f/ff/2015-Glasgow-interlab21.png" height="70%" width="70%"/> | ||
<figcaption>Table 6: Derived measurements of devices and controls.</figcaption> | <figcaption>Table 6: Derived measurements of devices and controls.</figcaption> | ||
</br> | </br> | ||
Line 633: | Line 646: | ||
We attempted to estimate the absolute number of GFP molecules per cell (Table 7) using our phiLOV results and some simplifying assumptions. | We attempted to estimate the absolute number of GFP molecules per cell (Table 7) using our phiLOV results and some simplifying assumptions. | ||
</br> | </br> | ||
− | <img src="https://static.igem.org/mediawiki/2015/9/98/2015-Glasgow-Interlab24.png" height="70%" width="70%"/> | + | </figure></center> |
+ | <br> | ||
+ | <center><figure> <img src="https://static.igem.org/mediawiki/2015/9/98/2015-Glasgow-Interlab24.png" height="70%" width="70%"/> | ||
<figcaption> Table 7: Summary of absolute number of GFP molecules per cell.</figcaption> | <figcaption> Table 7: Summary of absolute number of GFP molecules per cell.</figcaption> | ||
In order to estimate the absolute number of GFP molecules per cell the following calculations were carried out: | In order to estimate the absolute number of GFP molecules per cell the following calculations were carried out: |
Revision as of 12:22, 18 September 2015
Interlab Study
Home > Measurement > Interlab Study
Overview
All 2015 iGEM teams have been invited to participate in the Second International InterLab Measurement Study in synthetic biology. Each lab will obtain fluorescence data for the same three GFP-coding devices with different promoters varying in strength. The objective is to assess the robustness of standard parts and the variability of measurements among different research groups using different lab techniques.
Introduction
This year iGEM Glasgow have participated in the InterLab study and Extra Credit. The three devices required were cloned, as specified, and using a plate reader measurements were obtained in absolute units in terms of moles of FAM labelled oligonucleotide.
Release
Individuals responsible for conducting InterLab study
Others who should be credited, e.g., in a publication based on this data
Dates of InterLab Study
Detailed Lab Book
Equipment
Equipment used to acquire measurements
o Incubator – 2cm shaking diameter o Life Science Analyzer: BioMate™ 3S Spectrophotometer – Used to measure absorbance at 600nm of each sample. o Typhoon FLA 9500 - GE Healthcare Life Sciences. Wavelength used to excite cells - 475nm. Filter/channel used to capture the light emission from the cells - Filter BPB1 (530DF20).
Spectrophotometer calibration
Typhoon FLA 9500 calibration
Methodology
Protocol for cloning devices
Preparation for measurements
Protocol for measurements
The controls
Protocol for calculating a conversion factor for absolute units
Measurements
Direct Measurement (Raw Data)
Derived Measurements (Conversion to Absolute units)
Estimation of absolute number of GFP molecules per cell
References
Buckley, A. Petersen, J. Roe, A. Douce, G. Christie, J. (2015). LOV-based reporters for fluorescence imaging. Current Opinion in Chemical Biology. 27 (1), p39–45.