− | One of the vital elements of a Do-It-Yourself bioreactor is the hardware design which should be feasible as well as reliable. The most important point to be kept in mind while constructing a bioreactor is the sterility. At every step of the development of the bioreactor, we have fulfilled | + | One of the vital elements of a Do-It-Yourself bioreactor is the hardware design which should be feasible as well as reliable. The most important point to be kept in mind while constructing a bioreactor is the sterility. At every step of the development of the bioreactor, we have fulfilled these criteria. |
{{Team:Aachen/ReadMore|title=Construction Manuals|link=/Team:Aachen/Notebook/Construction_Manuals|picture=rmConstruction|url=/wiki/images/1/15/Aachen_tile_Lab_Bioreactor_Hardware_construction_manuals.JPG}} | {{Team:Aachen/ReadMore|title=Construction Manuals|link=/Team:Aachen/Notebook/Construction_Manuals|picture=rmConstruction|url=/wiki/images/1/15/Aachen_tile_Lab_Bioreactor_Hardware_construction_manuals.JPG}} | ||
Line 23: | Line 23: | ||
==Principle== | ==Principle== | ||
− | For our bioreactor, we need a peristaltic pump and it is necessary to use sterile tubings which requires the pumps to be assembled and disassembled in minimum possible time. Hence the pumps were built with different layers and a 3D structure that could be partly disassembled to position the tube. | + | For our bioreactor, we need a peristaltic pump and it is necessary to use sterile tubings which requires the pumps to be assembled and disassembled in minimum possible time. Hence the pumps were built with different layers and a 3D structure that could be partly disassembled to position the tube. The tube was pressed between the 3D structure and the walls thereby generated the required pressure to pump the liquid from one side to another. |
A detailed construction manual is available at '''[[Team:Aachen/Notebook/Construction_Manuals/Pumps|pumps]]'''. | A detailed construction manual is available at '''[[Team:Aachen/Notebook/Construction_Manuals/Pumps|pumps]]'''. | ||
=Stirrer= | =Stirrer= | ||
− | {{Team:Aachen/Figure|Aachen_Reactor_Vortex.JPG|title=The vortex inside the reactor | + | {{Team:Aachen/Figure|Aachen_Reactor_Vortex.JPG|title=The vortex inside the reactor created by our stirrer|subtitle=The stirring range is between 200 rpm and 1400 rpm|size=medium}} |
We use a DC motor and a 3D printed structure holding 4 magnets as our stirrer. The Voltage to the circuit is controlled using Analog output values thus controlling the speed of the stirrer. The magnets are placed in such a way so that half of the structure is north and the other half is south. This increases the area of magnetism thus holding the stirrer firmly into its magnetic pull. | We use a DC motor and a 3D printed structure holding 4 magnets as our stirrer. The Voltage to the circuit is controlled using Analog output values thus controlling the speed of the stirrer. The magnets are placed in such a way so that half of the structure is north and the other half is south. This increases the area of magnetism thus holding the stirrer firmly into its magnetic pull. | ||
Line 45: | Line 45: | ||
− | Light having wavelength 605 nm is emitted by a LED and is emitted in the medium. On the opposite side, we have a | + | Light having wavelength 605 nm is emitted by a LED and is emitted in the medium. On the opposite side, we have a phototransistor to measure the intensity of the light reaching it, through the sample. |
− | We have developed on iGEM Aachen 2014 OD device measurement principle to get live data, and built a device using optimum materials, along with use of calibration file to get as precise | + | We have developed on iGEM Aachen 2014 OD device measurement principle to get live data, and built a device using optimum materials, along with use of calibration file to get as precise values as possible. |
To obtain the cell concentration in the medium, a special transistor that converts light to frequency, TSL 235R<ref>http://www.ti.com/lit/ds/symlink/tsl235.pdf</ref>, is used. | To obtain the cell concentration in the medium, a special transistor that converts light to frequency, TSL 235R<ref>http://www.ti.com/lit/ds/symlink/tsl235.pdf</ref>, is used. |