=Pumps=
We built upon a design from thingyverse <ref>http://www.thingiverse.com/thing:642192</ref> that uses a 4-wired NEMA 17 stepper motor with 1.8&nbsp;°/step to construct a peristaltic pump. The original design however did not fit our requirement of sterile assembly, because the tubing was pushed through a hole in a 3D-printed structure. Using Autodesk Inventor, we designed a different geometry to accomodate for rapid prototyping and sterile assembly.
+
We built upon a design from thingyverse <ref>http://www.thingiverse.com/thing:642192</ref> that uses a 4-wired NEMA 17 stepper motor with 1.8&nbsp;°/step to construct a peristaltic pump. The original design however did not fit our requirement of sterile assembly, because the tubing was pushed through a hole in a 3D-printed structure. Using Autodesk Inventor, we designed a different structure to accomodate rapid prototyping and sterile assembly.
 
{{Team:Aachen/Figure|Aachen_PumpConstruct7.jpg|title=DIY Pump|subtitle=Fully Assembled Pump in Motion.|size=large}}
 
{{Team:Aachen/Figure|Aachen_PumpConstruct7.jpg|title=DIY Pump|subtitle=Fully Assembled Pump in Motion.|size=large}}
  
 
==Principle==
 
==Principle==
For our bioreactor, we need a peristaltic pump and it is necessary to use sterile tubings. This can be achieved if the pumps can be assembled and disassembled quickly. Hence the pumps were built with different layers and a 3D structure that can be partly disassembled to position the tube. The tube gets pressed between the 3D structure and the walls thereby generating the required pressure to pump the liquid from one side to another.
+
For our bioreactor, we need a peristaltic pump and it is necessary to use sterile tubings which requires the pumps to be assembled and disassembled in minimum possible time. Hence the pumps were built with different layers and a 3D structure that could be partly disassembled to position the tube. ??The tube was pressed between the 3D structure and the walls thereby generated the required pressure to pump the liquid from one side to another.??
  
 
A detailed construction manual is available at '''[[Team:Aachen/Notebook/Construction_Manuals/Pumps|pumps]]'''.
 
A detailed construction manual is available at '''[[Team:Aachen/Notebook/Construction_Manuals/Pumps|pumps]]'''.

Revision as of 22:55, 18 September 2015


Overview

One of the vital elements of a Do-It-Yourself bioreactor is the hardware design which should be feasible as well as reliable. The most important point to be kept in mind while constructing a bioreactor is the sterility. At every step of the development of the bioreactor, we have fulfilled this criteria.

Key Achievements

  • design and characterization of DIY peristaltic pumps with a dynamic range of 0.5-1400 ml/h
  • development of a semi-continuous OD measurement unit
  • calibrated magnetic stirring from 100-1400 rpm
  • integrated circuit for controlling three pumps and stirring
  • designed DIY constructions to hold circuit, pumps and bioreactor setup
  • parallelization of multiple reactor control units
  • characterization of DIY chemostat bioreactors

Pumps

We built upon a design from thingyverse [1] that uses a 4-wired NEMA 17 stepper motor with 1.8 °/step to construct a peristaltic pump. The original design however did not fit our requirement of sterile assembly, because the tubing was pushed through a hole in a 3D-printed structure. Using Autodesk Inventor, we designed a different structure to accomodate rapid prototyping and sterile assembly.

Aachen PumpConstruct7.jpg
DIY Pump
Fully Assembled Pump in Motion.

Principle

For our bioreactor, we need a peristaltic pump and it is necessary to use sterile tubings which requires the pumps to be assembled and disassembled in minimum possible time. Hence the pumps were built with different layers and a 3D structure that could be partly disassembled to position the tube. ??The tube was pressed between the 3D structure and the walls thereby generated the required pressure to pump the liquid from one side to another.??

A detailed construction manual is available at pumps.

Stirrer

Aachen Reactor Vortex.JPG
The vortex inside the reactor creator by our stirrer
The stirring range is between 200 rpm and 1200 rpm

We use a DC motor and a 3D printed structure holding 4 magnets as our stirrer. The Voltage to the circuit is controlled using Analog output values thus controlling the speed of the stirrer. The magnets are placed in such a way so that half of the structure is north and the other half is south. This increases the area of magnetism thus holding the stirrer firmly into its magnetic pull. Contruction Manual can be found on stirrer.

Continous OD Device

How can we find out if there is really bacteria growing in our reactor? The common way is doing OD measurements. But if there is only 10 ml of reactor volume, every measurement would disturb the conditions drastically and thereby influence the experiment itself.

Aachen ODCloseup.jpg
Optical Density Measurement device
As shown the transmitter and receiver are placed opposite sides of the tube


It is most important that the samples are taken sterile. Furthermore the OD should be measured automatically so that the experiment can run without the need of a person taking samples. Finally the device to measure to OD has to be cheap and easy to assemble while having a precision to compete with commercial spectrometers.


Measurement Principle

Light with a wavelength of 605 nm is emitted by a LED and shines through the medium. On the other side we have a phototransister to measure the intensity of the light reaching through the sample, therefore we use a TSL 235 R, a special transitor converting light to frequency. We have developed on iGEM Aachen 2014 OD device measurement principle to get live data, and built using optimum materials, along with use of calibration file to get as precise value as possible.

To keep the whole measurement sterile, the medium with the bacteria flows through silicon tubes and a transparent glass structure through the device without getting in contact with the surrounding air, while the device itself is mounted around this tube.


A detailed construction manual is available at biomass sensor.


To test if our continuous OD device can also be used to monitor batch fermentations, we set up a simple batch culture. We used a preculture of #RP1O# and M9 medium with 40 mM glucose that was grown in a stirred flask. The continuous OD sensor was attached right above the flask and a pump was used to cycle culture from the vessel through the sensor and back into the vessel. This setup was far away from perfect from a bioprocess perspective, but the goal of this experiment was testing the OD sensor and not the generation of biologically relevant samples.

Aachen OnlineOD.png
Growth Curve Recorded with our Online OD Unit
The growth rate is smoothed with a moving average over 150 data points.

Assembly

The assembly of the bioreactor electronics was the biggest challenge that we faced. Each device had a pin protocol which was maintained and tagged to the controling firmware. The assembled final firmware had to be checked and rechecked to compromise between each of the seperated devices. Finally we came up with idea of not using delays and using counters in our firmware to make it look easy and compatible to all components.

Outlook

Since in the current version of our project all digital ports of the Arduino Uno are in use, for future designs it will be necessary to change to a bigger controller, for example an Arduino Mega.


Then it will be possible to include further analysis units. We allready took some steps towards an offgas analysis with CO2 and O2 sensors, and CHX analysis would be interesting for many experiments. An integration of these components would be very easy. Even more we cooperated with a local community lab, the Technik Garage, who developed a pumping and heating system for the water bath for us. To include this in the whole system had given us much more parameters to control the experiment.


Biological experiments often run far more than a week. It can be very time-consuming to have a look at the experiment al the time. Therefore it would be comfortable to control your experiments via a local net or even the Internet. Imagine lying on a meadow in summer with a cool drink at hand and checking the OD instead of going to the lab every time.


It is obvious that there are many challenges to work on with even more opportunities. Hence the modularity and open source features of our system serve as a platform for future developments.

References

  1. http://www.thingiverse.com/thing:642192