<p>Our objective was therefore twofold: firstly to test the capabilities of OpenScope in analysing bacterial samples for GFP and RFP fluorescence, and secondly to report back to Glasgow with our phenotypic analysis of their bacterial strains.</p> | <p>Our objective was therefore twofold: firstly to test the capabilities of OpenScope in analysing bacterial samples for GFP and RFP fluorescence, and secondly to report back to Glasgow with our phenotypic analysis of their bacterial strains.</p> | ||
− | |||
<h4> Results:</h4> | <h4> Results:</h4> | ||
− | |||
<p>After preliminary testing using the RFP epi-cube, it was decided that imaging of RFP at this stage would not be possible. Hence only bacterial strains expressing GFP (as confirmed earlier) were tested against the control.</p> | <p>After preliminary testing using the RFP epi-cube, it was decided that imaging of RFP at this stage would not be possible. Hence only bacterial strains expressing GFP (as confirmed earlier) were tested against the control.</p> | ||
<p>Earlier testing using the standard set-up (single LED illumination, GFP epi-cube) indicated that visualising fluorescent beads was possible with OpenScope (Fig. 1). </p> | <p>Earlier testing using the standard set-up (single LED illumination, GFP epi-cube) indicated that visualising fluorescent beads was possible with OpenScope (Fig. 1). </p> | ||
Line 37: | Line 35: | ||
<div style="float:left; margin-right: 20px"><p><i><b>Fig. 3:</b> <b>a</b> Control slide (left) showing no fluorescence <b>b</b> p126.1 samples imaged using non-standard set-up under 470nm excitation (centre left)<b>c</b> p126.1+p80.1 samples imaged using non-standard set-up showing fluorescence (centre right) <b>d</b>p126.1+p56.1 samples imaged using non-standard set-up (right). All images were captured using the Webshell on 15.09.15 and are unedited.</i></p></div> | <div style="float:left; margin-right: 20px"><p><i><b>Fig. 3:</b> <b>a</b> Control slide (left) showing no fluorescence <b>b</b> p126.1 samples imaged using non-standard set-up under 470nm excitation (centre left)<b>c</b> p126.1+p80.1 samples imaged using non-standard set-up showing fluorescence (centre right) <b>d</b>p126.1+p56.1 samples imaged using non-standard set-up (right). All images were captured using the Webshell on 15.09.15 and are unedited.</i></p></div> | ||
<br style="clear: both"> | <br style="clear: both"> | ||
+ | </div></div></section> | ||
<br> | <br> | ||
Line 87: | Line 86: | ||
<br> | <br> | ||
<h4> Results:</h4> | <h4> Results:</h4> | ||
− | |||
<p>Testing using the commercial fluorescence microscope confirmed that the samples had phenotypes as reported by William and Mary. Both constructs resulted in GFP expression in E. coli. J23106+I13504 demonstrated increased fluorescence intensity relative to J23117+I13504.</p> | <p>Testing using the commercial fluorescence microscope confirmed that the samples had phenotypes as reported by William and Mary. Both constructs resulted in GFP expression in E. coli. J23106+I13504 demonstrated increased fluorescence intensity relative to J23117+I13504.</p> | ||
<p> A control slide (sample taken from the agar of the control plate, untransformed cells plated on Amp) was first tested to establish a fluorescence-free baseline (Fig. 2a). The duplicate J23106+I13504 samples were then tested using the standard set-up (100mW LED) and non-standard set-up (3W LED). Fluorescence was detected (Fig. 2b and 2c).</p> | <p> A control slide (sample taken from the agar of the control plate, untransformed cells plated on Amp) was first tested to establish a fluorescence-free baseline (Fig. 2a). The duplicate J23106+I13504 samples were then tested using the standard set-up (100mW LED) and non-standard set-up (3W LED). Fluorescence was detected (Fig. 2b and 2c).</p> |