Difference between revisions of "Team:Cambridge-JIC/Collaborations"
Maoenglish (Talk | contribs) |
Maoenglish (Talk | contribs) |
||
Line 72: | Line 72: | ||
<h4> Results:</h4> | <h4> Results:</h4> | ||
<h4> Conclusions:</h4> | <h4> Conclusions:</h4> | ||
+ | <p>The first objective of the collaboration was to confirm the phenotypes of the bacterial strains and the expression of fluorescent proteins. Due to the technical limitations of OpenScope (see below), RFP could not be visualised. Hence a commercially available microscope was used to confirm RFP expression. In addition, GFP was reliably confirmed using the same microscope. Across the board, expression of the fluorescent proteins was as reported by Glasgow and W&M iGEM teams. This confirms the functionality of the plasmids. </p> | ||
+ | <p>After successful visualisation of fluorescent beads labeled with GFP (Fig. 1), it was expected that OpenScope would enable visualisation of E. coli expressing GFP. However, results indicate that reliable detection was not possible using the standard set-up (single LED illumination, GFP epi-cube). Possible explanations for this are as follows:</p> | ||
+ | <ol> | ||
+ | <li><p>The fluorescent beads used are far brighter than biological samples, and therefore can be visualised with low-brightness LEDs</li></p> | ||
+ | <li><p>The bacteria, when prepared on slides, form a relatively uniform thin surface. Flourescence from this is difficulty to detect, as OpenScope cannot resolve individual bacterial cells</li></p> | ||
+ | </ol> | ||
+ | <p>Replacement of the 100mW LED with the 3W LED allowed visualisation of samples with reduced fluorescence intensity in the case of the p126.1, p126.+p56.1 and p126.1+p80.1 cells. The 100mW LED was sufficient only to image the J23106+I13504 samples. Overall, the results suggest that in order to reliably detect fluorescence the 3W LED is more appropriate. However, this is still not sufficiently reliable to make OpenScope useful for fluorescence screening at this stage. In addition, the artifact (image of the LED itself) seen when using the 3W LED means that the uniformity of illumination must be improved.</p> | ||
+ | <p>RFP is more challenging to image, as it has a narrow gap between the excitation (584nm) and emission (607nm). Hence it is difficult to find low-cost dichroic mirrors that are transparent to wavelengths around 607 nm while being reflective to wavelengths around 584 nm. In addition, sourcing LEDs with an emission peak in the region of 584 nm was not possible. As such, the LEDs used had an emission peak at 591nm, which is closer to the transparency region for the dichroic mirrors. Overall, RFP imaging has not yet been demonstrated as a proof of concept. Perhaps this could be starting point for future iGEM teams looking to build on the OpenScope project. </p> | ||
</div></div></section> | </div></div></section> | ||
Revision as of 22:25, 15 September 2015