− | < | + | <h3> Results:</h3> |
<p>After preliminary testing using the RFP epi-cube, it was decided that imaging of RFP at this stage would not be possible. Hence only bacterial strains expressing GFP (as confirmed earlier) were tested against the control.</p> | <p>After preliminary testing using the RFP epi-cube, it was decided that imaging of RFP at this stage would not be possible. Hence only bacterial strains expressing GFP (as confirmed earlier) were tested against the control.</p> | ||
<p>Earlier testing using the standard set-up (single LED illumination, GFP epi-cube) indicated that visualising fluorescent beads was possible with OpenScope (Fig. 1). </p> | <p>Earlier testing using the standard set-up (single LED illumination, GFP epi-cube) indicated that visualising fluorescent beads was possible with OpenScope (Fig. 1). </p> | ||
Line 80: | Line 80: | ||
<p>Testing using the commercial fluorescence microscope confirmed that the samples had phenotypes as reported by Glasgow. From tube 5, a small proportion of the cells were expressing RFP and the majority expressed GFP as predicted.</p> | <p>Testing using the commercial fluorescence microscope confirmed that the samples had phenotypes as reported by Glasgow. From tube 5, a small proportion of the cells were expressing RFP and the majority expressed GFP as predicted.</p> | ||
<p>Results from preliminary testing of DH5α cells with p126.1 and p56.1 (confirmed: GFP expression only) using the standard set-up indicated illumination brightness was insufficient to detect GFP. The non-standard set-up was used, and GFP expression was confirmed in p126.1, p126.+p56.1 and p126.1+p80.1 cells as expected (Fig. 3a-d). The images display an artefact of the square shape of the LED used, as there is an area of increased fluorescence in the outline of a square at the centre of the images (Fig 3c and d). </p> | <p>Results from preliminary testing of DH5α cells with p126.1 and p56.1 (confirmed: GFP expression only) using the standard set-up indicated illumination brightness was insufficient to detect GFP. The non-standard set-up was used, and GFP expression was confirmed in p126.1, p126.+p56.1 and p126.1+p80.1 cells as expected (Fig. 3a-d). The images display an artefact of the square shape of the LED used, as there is an area of increased fluorescence in the outline of a square at the centre of the images (Fig 3c and d). </p> | ||
− | < | + | <br> |
+ | <h3> Conclusions:</h3> | ||
<p>The first objective of the collaboration was to confirm the phenotypes of the bacterial strains and the expression of fluorescent proteins. Due to the technical limitations of OpenScope (see below), RFP could not be visualised. Hence a commercially available microscope was used to confirm RFP expression. In addition, GFP was reliably confirmed using the same microscope. Across the board, expression of the fluorescent proteins was as reported by Glasgow and W&M iGEM teams. This confirms the functionality of the plasmids. </p> | <p>The first objective of the collaboration was to confirm the phenotypes of the bacterial strains and the expression of fluorescent proteins. Due to the technical limitations of OpenScope (see below), RFP could not be visualised. Hence a commercially available microscope was used to confirm RFP expression. In addition, GFP was reliably confirmed using the same microscope. Across the board, expression of the fluorescent proteins was as reported by Glasgow and W&M iGEM teams. This confirms the functionality of the plasmids. </p> | ||
<p>After successful visualisation of fluorescent beads labeled with GFP (Fig. 1), it was expected that OpenScope would enable visualisation of E. coli expressing GFP. However, results indicate that reliable detection was not possible using the standard set-up (single LED illumination, GFP epi-cube). Possible explanations for this are as follows:</p> | <p>After successful visualisation of fluorescent beads labeled with GFP (Fig. 1), it was expected that OpenScope would enable visualisation of E. coli expressing GFP. However, results indicate that reliable detection was not possible using the standard set-up (single LED illumination, GFP epi-cube). Possible explanations for this are as follows:</p> |